DO Probe
- Electrochemical Method:
Composed of two metal electrodes in contact with supporting electrolyte and separated from the test solution by a gas permeable membrane.
A constant voltage is placed across the cathode and anode. Oxygen diffuses through the membrane and is reduced at the cathode by the voltage.
This process produces a current flow, which is detected by the meter and is proportional to the partial pressure of oxygen.
- Saves money, time, and labor
- No preparation of reagents or titration
- Allows for continuous measurement.
Polarographic: Single vs. Dual Thermistor
- Added thermistor to electrolyte (in addition to sample sensor).
This allows temperature measurement in sample [water or air] AND in electrolyte.
Air is poor heat sink (think of cooking in a copper vs. aluminum pot).
- People have tendency to calibrate too quickly when doing AIR calibrations.
- Double thermistor monitors differential between air and electrolyte and does not lock in calibration until the two are equal in temperature.
- Bottom Line: provides for more accurate & consistent calibrations.
Back to Top
Luminescence
A whole new technology for environmental chemistry, but which has been in use in medical field for years
Luminescence technology I: Initially Blue light from an LED inside the probe excites luminerscent material on the sensor surface.
Image courtesy of HACH, Inc.
Probes utilize a sensor coated with luminescent material
Blue light is transmitted to the sensor from an LED on the surface
This blue light excites the luminescent material which in turn emits red light as it returns to its unexcited state
The elapsed time from excitation till return to steady state is measured.
The more oxygen present, the shorter the time it takes for red light emission.
Time is measured and correlated to oxygen concentration.
Luminescence technology II: As oxygen molecules encounter the sensor surface, they react with the luminescent material, causing it to return to its un-excited state. This results in emission of red light which is correlated to oxygen content.
Image courtesy of HACH, Inc.
- Developed by the HACH company
- Probe uses NO:
- Anode
- Electrolyte
- Cathode
- Membrane
- All of the items listed above are potential sources of error in conventional DO technology
- Research shows luminescence requires
- Less frequent adjustments during calibration
- Less maintenance
- Dis-advantages of luminescence probe
- Requires an awkward funnel and magnetic stirrer
- The funnel is difficult to use and poses a cross-contamination risk
Back to Top
- YES!
- Changes to ch. NR 219, Wis,. Admin. Code, which took effect June 1, 2009 included
approved methodologies for Luminescence technology.
- Electrolyte replenishment
- Membrane failure
- Membrane rupture
- Membrane fouling
- Cathode and anode cleaning
- Follow manufacturer recommendations for interval & procedure
Troubleshooting DO Probe Problems:
see NCL website (www.nclabs.com) for solid information regarding diagnosing DO probe problems.
Some information included at the end of this presentation.
- DO µA (micro-amps) should be from 8.0 to 17.0
- %/ µA (micro-amps) should be from 5.9 to 12.6%
- Replace membrane if outside this range
.
- Good tool for preventative maintenance
YSI on-board tools for probe maintenance: Step 1- Press 'Diagnosis'
>YSI on-board tools for probe maintenance: Step 1- Press 'Diagnosis'
YSI on-board tools for probe maintenance: Step 3 - The display will now show probe values for microamps(uA) and microAmp percentage (uA %).
Back to Top
- Examine membrane daily
- Replace membrane if air bubbles, wrinkles, or if there is buildup on or under the membrane
- If gold tip (cathode) is tarnished, clean it
- The gold should have a bright "matte" finish with fine scratches-DO NOT polish too much!
- Clean with sanding disk provided in the YSI cap
- The silver anode should have a light silver color
- Clean if anode is dark
- Clean by soaking overnight in household ammonia solution
- Rinse thoroughly with tap water, DI water and electrolyte
- Poor calibration may give the appearance of a dilution water problem when the water may be fine
- Recommend calibrating using the air-saturated WATER calibration or Winkler titration
- If air-saturated water calibration is used, use a good quality barometer in the laboratory
- Check the barometer calibration against a reliable source at least quarterly (internet, airport, local station). Remember you must re-correct for actual altitude.
Back to Top
Information obtained from www.nclabs.com
- Allow ≥ 2 hr after membrane change for the probe to stabilize. Overnight is better.
- Warm-up instrument. Calibrate.
- Observe readings continuously for 2 mins. w/probe in bottle.
- Be sure the temperature is constant.
- Watch the readings carefully.
DO NOT just record the initial reading and come back 2 minutes later.
You need to actually see what happens over the time period.
- If readings drifts slowly DOWN, a longer warm up time is required.
- If readings JUMP AROUND, the probe is not functioning properly.
- If readings STABLE in the air calibration bottle, sensor is probably OK.
- If readings stable in the air calibration bottle but not in solution, the membrane is probably defective.
Zero Oxygen Check (Response check):
- Dissolve 0.5-1 grams of sodium sulfite (Na2SO3) in 300 mL of lab reagent water.
- Stir slowly--avoid “tornadoes”; slowly pour into a BOD bottle.
- Calibrate your DO probe as you normally would.
- Place the probe into the "Zero Oxygen" solution
- Observe!
- Meter should read "0" within two minutes.
- (With some older YSI systems, readings below 1.0 mg/l are considered zero.)
Be sure to rinse everything thoroughly. Sodium sulfite is a significant oxygen scavenger, so residue left behind will result in oxygen depletion from samples.
Back to Top
Last Revised: Monday March 08 2010
|